The TA-1200P Thermocouple Furnace performs functions that would require three different types of instruments: a furnace, standard thermometer and electronic calibrator measuring signals generated by RTDs, thermocouples, thermoswitches and mA generating sources. Besides these functions it has HART® Communication Protocol, data logger mode, Ethernet and USB interface.
- Generates temperatures from 50 ºC up to 1200 ºC with a resolution of 0.1 ºC.
- Allows the use of a black body insert for the calibration of infrared pyrometers.
- Using the internal reference, accuracy of up to ±2.20 ºC.
- Using an external thermometer such as Presys Super Thermometer ST-501, accuracy of up to ±0.75 ºC.
- Stability of ±0.1 ºC.
- Uniformity of ±0.2 ºC.
- Input for RTD, thermocouples and thermoswitches.
- Internal regulated 24 Vdc power supply and mA input for 2-wire transmitters.
- External reference thermometer is optional.
- Complete automatic calibration without the use of a computer.
- Advanced Documenting capabilities: connection with computer and ISOPLAN calibration software.
- Portable, compact, with carrying case and interchangeable inserts.